After preparing the buffer solution, it should be placed in a glass bottle or a polyethylene bottle (alkaline pH buffer solutions such as pH 9.18, pH 10.01, pH 12.46, etc. should be placed in polyethylene bottles). The bottle cap should be tightly closed and stored at low temperature (5-10 ℃) in the refrigerator. It can generally be used for about six months. If turbidity, mold or precipitation are found, it cannot be continued to be used. When using, prepare several 50mL polyethylene small bottles, pour the buffer solution from the large bottles into the small bottles, and let them stand at ambient temperature for 1-2 hours until the temperature is balanced before use. After use, it should not be poured into the large bottle to avoid contamination. The buffer solution in the bottle can be used for 2-3 days under environmental conditions of>10 gráður. Almennt er hægt að nota pH 7,00, pH 6,86 og pH 14,00 lausnir í lengri tíma. pH 9,18 og pH 10,01 lausnir eru viðkvæmar fyrir pH-breytingum vegna frásogs koltvísýrings úr loftinu.
Hver er tilgangurinn með pH buffer lausn?
Kvarðaðu pH mælinn áður en þú mælir pH.
Til að sannreyna nákvæmni pH-mælis, til dæmis, eftir að hafa kvarðað pH-mæli með pH 6,86 og pH 14,00, skal setja pH rafskautið í pH 9,18 lausnina og athuga hvort skjágildi tækisins sé í samræmi við pH gildi staðallausnarinnar.
Er nauðsynlegt að endurkvarða pH mælinn við almenna nákvæmnismælingu. Eftir að hafa kvarðað og notað pH-mælirinn getur rek eða breytingar átt sér stað. Þess vegna, áður en þú prófar, skaltu setja rafskautið í staðlaða biðminni lausn sem er nálægt mældu lausninni og ákvarða hvort endurkvörðun sé nauðsynleg miðað við stærð villunnar.